上海榮計達儀器科技有限公司
參 考 價 | 面議 |
產(chǎn)品型號SHBY-30
品 牌
廠商性質其他
所 在 地上海市
紡織服裝機械網(wǎng)采購部電話:0571-88918531QQ:2568841715
聯(lián)系方式:查看聯(lián)系方式
更新時間:2023-10-18 12:34:05瀏覽次數(shù):72次
聯(lián)系我時,請告知來自 紡織服裝機械網(wǎng)石膏恒溫恒濕箱是根據(jù)國家A新標準(石膏、砂漿,涂料,膨脹劑,保溫材料,節(jié)能材料)養(yǎng)護設計而成.操作簡單,全自動控制,可在任意氣溫和水溫條件下,控制溫度20℃±1,20℃-25℃之間進行試體養(yǎng)護,濕度控制在50%-80%之間
石膏恒溫恒濕箱是石膏檢測儀器,根據(jù)國家A新標準(石膏、砂漿,涂料,膨脹劑,保溫材料,節(jié)能材料)養(yǎng)護設計而成.操作簡單,全自動控制,可在任意氣溫和水溫條件下,控制溫度20℃±1,20℃-25℃之間進行試體養(yǎng)護,濕度控制在50%-80%之間,使用前務必仔細閱讀說明書。并由**實驗人員操作,以避免操作不當引起的傷害。
The gypsum standard curing box is a gypsum testing instrument, designed according to the new national A standard (gypsum, mortar, coatings, expansion agents, thermal insulation materials, energy-saving materials). It is simple to operate, fully automatic control, and can be used at any temperature and water temperature. Under the conditions, the temperature is controlled between 20 ℃ ± 1, 20 ℃ -25 ℃, and the humidity is controlled between 50% -80%. Please read the instructions carefully before use. It is operated by ** experimental personnel to avoid injury caused by improper operation.
石膏恒溫恒濕箱技術參數(shù):
★電源: 220V
★箱內尺寸: 550× 550×1100(mm)
★恒溫范圍: 0--50℃可調
★恒濕范圍: 35%--95%可調
★制冷功率: 145W
★加執(zhí)功率: 300W
★加濕器功率: 15W
★去濕功率: 245W
★外型尺寸: 1200×1550×650 (mm)
★重量:120kg
Technical parameters of gypsum standard curing box:
★ Power: 220V
★ Box size: 550 × 550 × 1100 (mm)
★ Constant temperature range: 0--50 ℃ adjustable
Constant humidity range: 35% -95% adjustable
★ Cooling power: 145W
★ Additional power: 300W
Humidifier power: 15W
★ Dehumidifying power: 245W
★ External dimensions: 1200 × 1550 × 650 (mm)
★ Weight: 120kg
石膏恒溫恒濕箱操作規(guī)程:
①應將儀器置于通風干燥處,在環(huán)境溫度2~38℃,相對濕度不大于85%,周圍無強烈振動及強電磁場影響的室內使用。
②砂漿養(yǎng)護箱安放平穩(wěn)后,應檢查門開關是否靈活,電氣、制冷等部件是否損壞。
③將箱內水箱加水至末層擱架以下,電熱管以上2cm,把側門打開將加濕器水箱內加注滿蒸餾水,另把測溫干濕盒內塑料球里注滿水,并把紗布泡入水中。
④將電源插頭插入具有良好接地的220V插座中,HBY-30按動電源開關,儀器自動進入工作狀態(tài)。
⑤出廠時砂漿養(yǎng)護箱已編入標準設定程序,溫度控制值為20.0℃±1.0℃,濕度控制值為95%±1.5%,無特殊要求,不能改變控制程序,否則將可能出現(xiàn)控溫、控濕超出技術范圍,如需要改變控制值數(shù)據(jù),可按照控溫儀說明書修改程序。
注意事項:
①電源電壓波動超過規(guī)定使用范圍,必須安裝電源穩(wěn)壓器,否則將會損壞儀表及制冷機。
②電源接地一定要可靠,接地不良會對您的人身安全帶來危險。
③經(jīng)常檢查加濕器、水箱,干濕盒內不要斷水,盒內紗布要每月更換。
④非有關人員不能擅自改變控制程序數(shù)據(jù),以免造成儀器失控及損壞儀器。
Gypsum standard maintenance box operating procedures:
① The instrument should be placed in a ventilated and dry place, and used indoors with an ambient temperature of 2 ~ 38 ° C and a relative humidity of not more than 85%, without strong vibration and strong electromagnetic fields.
② After the mortar curing box is placed stably, check whether the door switch is flexible and whether the electrical and refrigeration components are damaged.
③ Fill the water tank in the tank to below the last shelf, 2cm above the electric heating tube, open the side door to fill the water tank of the humidifier with distilled water, fill the plastic ball in the temperature measurement wet and dry box with water, and soak the gauze Into the water.
④ Insert the power plug into a 220V socket with a good ground. Press HBY-30 to switch on the power and the instrument will automatically enter the working state.
⑤ The mortar curing box has been programmed into the standard setting procedure when leaving the factory. The temperature control value is 20.0 ℃ ± 1.0 ℃, and the humidity control value is 95% ± 1.5%. There is no special requirement, and the control program cannot be changed, otherwise temperature control, Humidity control is beyond the technical range. If you need to change the control value data, you can modify the program according to the instruction of the temperature controller.
Precautions:
① The power supply voltage fluctuates beyond the specified use range, and a power regulator must be installed, otherwise the instrument and the refrigerator will be damaged.
② The power supply grounding must be reliable. Poor grounding will bring danger to your personal safety.
③ Check the humidifier and water tank frequently. Do not cut off the water in the wet and dry box. The gauze in the box should be replaced every month.
④ Non-relevant personnel cannot change the control program data without authorization, so as not to cause the instrument to lose control and damage the instrument.
石膏恒溫恒濕箱常見故障排除方法:
①整機不工作:檢查電源開關,保險絲是否損壞。
②不制冷:檢查儀表是否顯示制冷狀態(tài),檢查壓縮機過載保護器,起動器是否損壞,壓縮機能正常運轉而不制冷則可能是制冷劑泄漏。
③加濕器不加濕:檢查加濕器內小風機是否工作,加濕器底座內水位是否過高,加濕器安放是否傾斜過大,霧化換能器是否干凈。
④箱內溫差大:檢查風機是否工作,制冷指示燈亮而風機不工作,則風機損壞或保險絲壞。
⑤偶爾受外界電信號沖擊,儀表出現(xiàn)數(shù)顯失控,只需關閉電源開關或拔下電源插頭,再重新開機,即能正常工作。
Common troubleshooting methods for HBY-30 plaster standard curing box:
① The whole machine does not work: check whether the power switch and fuse are damaged.
② Uncooled: Check whether the meter shows the cooling status, check the compressor overload protector, whether the starter is damaged, and the compressor can run normally without cooling. It may be refrigerant leakage.
③ Humidifier does not humidify: Check whether the small fan in the humidifier works, whether the water level in the base of the humidifier is too high, whether the humidifier is placed too lean, and whether the atomizing transducer is clean.
④ Large temperature difference in the box: check whether the fan is working, if the cooling indicator is on and the fan is not working, the fan is damaged or the fuse is broken.
⑤ Occasionally due to the impact of external electrical signals, the digital display is out of control. Just turn off the power switch or unplug the power plug, and then turn it on again to work normally.
您感興趣的產(chǎn)品PRODUCTS YOU ARE INTERESTED IN
紡織服裝機械網(wǎng) 設計制作,未經(jīng)允許翻錄必究 .? ? ?
請輸入賬號
請輸入密碼
請輸驗證碼
請輸入你感興趣的產(chǎn)品
請簡單描述您的需求
請選擇省份
聯(lián)系方式
上海榮計達儀器科技有限公司